TGL-20M desktop high speed refrigerated centrifuge

1. Do not move the centrifuge or open the door when the centrifuge is in operation.

2. The table top where the centrifuge is placed should be firm and flat, and all four rubber feet should be in contact with the table top and evenly stressed to avoid vibration.

3. The dosing of the centrifuge tube should be weighed and balanced. If the dosing difference is too large, it will produce large vibration. At this time, the machine should be stopped and checked to make the dosing meet the requirements. The centrifuge test tubes must be placed in an even number and symmetrical.

4. If the centrifugal test tube ruptures during operation, it will cause greater vibration and should be stopped immediately.

5. The time interval between restarting after each shutdown should not be less than 5 minutes to prevent damage to the compressor due to stalling.

6. After each centrifugation is completed, the rotor must be taken out, otherwise it will be rusted on the shaft for a long time, and the rotor will not be taken out, causing the whole centrifuge to be scrapped.

7. The power supply must have a ground wire.